For 20 years the standard in acute slice
laboratories worldwide. Also used in Dermatology, Embryology, &
Pathology. |
There are three
Vibroslice models as follows:
- Model HA752 - manual tissue transport
- Model MA752 - motorized tissue transport
- Model ROMA752 - remotely operated motorized tissue transport and
blade height
VibrosliceTM
Specifications: |
 |
|
Model MA752 Shown with
optional accessories:
-
Model 752LS Cold Light
Source
-
Model 752IM Inspection
Microscope
-
Model 765LP Peltier
Tissue Bath
-
Model 752/2/AC
Autoclavable Tissue Bath
|
|
Blade Motion: |
Straight Line |
|
Oscillation Speed: |
0Hz to 70Hz by rotary dial knob |
|
Amplitude of Cut: |
1mm nominal, all models |
|
Blade Advance Speed:
HA752
MA752
ROMA752 |
Manual
Motorized from 0.08mm/sec to 1.6mm/sec
Motorized from 0.08mm/sec to 1.6mm/sec |
|
Bath Return Speed:
HA752
MA752
ROMA752 |
Manual
Fixed speed of 1.7mm/sec
Fixed speed of 1.7mm/sec |
|
Section Thickness:
HA752
MA752
ROMA752 |
Manual with 2 micron resolution
Manual with 2 micron resolution
Dual speed motorized with resolution of 1 micron |
|
Vertical Range: |
32mm |
|
Sectioning Range: |
From 1mm to 40mm |
|
Maximum Specimen
Size: |
30mm x
40mm |
|
Tissue Bath: |
55mm x 82mm x 31mm / 140ml,
standard, autoclavable or cooled |
|
Cutting Head: |
Removable Autoclavable as standard |
|
Blades: |
Ceramic 7550/1/C or Stainless Steel
7550/1/SS |
|
Observation: |
Inspection Microscope 10x - 40x with
zoom or Magnifying Glass 2x and 'Clearview' window |
|
Power Supply: |
230VAC 50Hz or 115VAC
60Hz |
|
Power Rating: |
HA752, 15W; MA752 and
ROMA752, 20W |
|
Microtome Accessories:
-
Campden's Precision
Blades in Zirconium
-
Ceramic and Stainless
Steel
-
Temperature Controlled
Peltier Tissue Baths
-
Cold Light Source
-
Magnifying Glass or
Inspection Microscope
-
Autoclavable Baths
-
Autoclavable Cutting Head
-
Tilting Tissue Mount
|
General Description:
-
Vibroslice is designed to
cut sections of fresh, unfixed tissue for in-vitro recordings,
tissue culture and microassay work.
-
Vibroslice can be used to
section fixed tissue without previously embedding or freezing.
-
Vibroslice will section
tissue such as brain stem and liver, maintaining the integrity of
surface.
-
Ultra-low advance speed
is now available on motorized models.
-
The speed of oscillation
has been increased.
-
New blade holder for both
ceramic and slim stainless steel blades is incorporated on all
Vibroslice models.
-
The blade oscillates at a
fixed lateral displacement of 1.0mm at a frequency set by the user.
-
All models are supplied
with:
- Removable Tissue Holder and Bath (standard model)
- 50 Stainless Steel and two
Ceramic Blades
-
The Removable Tissue
Holder and Bath allows tissue to be prepared and stored, ready for
later use and provides for easy cleaning which prevents carryover of
fixative between fixed and unfixed tissue.
-
Original concept of John
Jefferys, (University of Birmingham Medical School, UK).
Ref: J.G.R. Jeffreys. The
Vibroslice, a new vibrating-blade tissue slicer. J.Physiol. (Lond.)
234:2P, 1982. (Abstract)
|
|
The motorized tissue
transport Model MA752 |
| Control of
the speed of advance of the tissue to be sectioned is set on a rotary
dial and the user then presses a footswitch leaving both hands free
during operation. This model is recommended for use in cutting serial
sections for reconstruction or for producing multiple sections. Both
the HA752 and the MA752 are fitted with the Standard Tissue Bath and
Holder. Optionally, this can be replaced by Autoclavable Bath and/or
Tilting Tissue Holder. The standard cutting head can be removed for
autoclaving in applications where sterility is important. |
Model HA752 (left) and
Model MA752 (right)

|
|
The manual tissue
transport Model HA752
The manual advance of the section gives complete control during
sectioning. The HA752 is recommended for use in producing a small
number of sections or for use on tissues with differing mechanical
properties.
|
|
The remote motorized
Model ROMA752 |
| This
instrument was designed especially for work with hazardous tissues
such as those containing Prions or rabies infection. The cutting head,
tissue bath and associated drive mechanisms are operated remotely from
the control box. Thus the slicing portion of this instrument can be
placed inside a controlled environment such as a thin film isolator. A
tissue bath and cutting head capable of withstanding autoclaving at
140 0 C are included. |

Model ROMA752 |
|
View our other Tissue Sectioning/Slice Systems
Information:
|
| |
 |